122 
Instrument Messages
Condition Messages
149 8 Meas Error
151 9 Memory Error E
151 9 Memory Error;Shorten 
capture interval
E A shortage of free memory related to longer 
capture intervals has occurred. The 
measurement is aborted and all results return 
invalid values
153 10 I/O Error E
153 10 I/O Error; Ext Source needs 
IP Addr
E No IP address entered for external source and 
external LO control is ON.
155 11 Trig Error E
157 12 Invalid Data Status bit 
only
This is the “invalid data indicator”, same as the 
“*” in the upper right corner of the screen. It 
means that the on-screen annotation does not 
match the on-screen data, usually because a 
measurement is pending after a settings 
change. There is no message in the status line 
and nothing in the history queue, but there IS 
an on-screen indication and a status bit.
159 13 Settings Alert W
159 13 Settings Alert;LO may 
overload IF
W If the sweep type is Swept, the start frequency 
of the instrument is less than 10 MHz, and you 
put Swept IF Gain in Manual High, then a 
warning condition is generated and remains in 
effect as long as this condition exists.
In some older analyzers this was error 1109.
159 13 Settings Alert; Diff probe 
mismatch; <I|Q|I,Q>
W The attenuation values of the two probes on 
the I and/or Q channels differ by too much for a 
valid differential reading.
159 13 Settings Alert; Acquisition 
truncated
W In the Analog Demod mode, certain extreme 
settings combinations will result in a required 
acquisition length in excess of the capacity of 
the analyzer.Increase the AF Spectrum RBW 
or the RF Spectrum RBW, decrease the 
Channel BW, and/or decrease the Demod 
Waveform Sweep Time.
Err# Bit in 
status 
register
Message Error or 
Warning
More Information