Assembly Replacement Procedures
RF Area
Figure 14-24 Installing New Coax Cables
4. Install W3 between J2 of the A13 Front End Assembly and J2 of the A11
Low Band Switch. Torque to 10 inch-pounds.
5. Install W10 between the Type N input connector and the input of the A9
Input Attenuator. Torque to 10 inch-pounds.
6. Refer to Figure 14-25. Place one of the two coaxial switches (1) and the
Switch 1 Bracket (2) onto the low band switch bracket. Mount the switch
with the Keysight label facing up. Attach the bracket and switch to the low
band switch bracket using two 0515-1992 screws (3) using the T-8 driver.
Do not torque the screws at this time