EasyManua.ls Logo

Keysight N9020A - Page 475

Keysight N9020A
642 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
475
Assembly Replacement Procedures
RF Area
Replacement
1. Refer to Figure 14-31. Position the attenuator in the bracket so that the
ribbon connector end is “up”.
2. Replace the two screws that attach the attenuator to the bracket. Torque
to 9 inch-pounds.
3. Position the attenuator and bracket into the chassis and attach with the
two screws. Torque to 9 inch-pounds.
4. Replace the ribbon cable and semi-rigid cables W9, W10, W11 and W19
to the attenuator. Torque the semi-rigid cables to 10 inch-pounds.