566 
Post-Repair Procedures
Post-Repair Procedures
NOTE Some Adjustment tests may be found in the Utility portion of the Test Plan.
W24 
BBIQ Cable 
Assembly
(Option BBA)
Press 
Input/Output, I/Q, I/Q Cable 
Calibrate
Absolute Amplitude (Option BBA) Adjust
Residual Responses
Displayed Average Noise Level
Absolute Amplitude Accuracy at 250 kHz
Frequency Response Below 250 kHz
Frequency Response Above 250 kHz
Channel Match
a. Calibration Application Software is required for all performance tests and most adjustments. For 
details go to http://www.keysight.com/find/calibrationsoftware 
b. IF Frequency Response adjustment only performed if Option B25 and/or B40 are present and ana-
lyzer is equipped with Option MPB.
c. See the Additional Tasks section in this chapter for detailed instructions on this task.
d. To set the instrument time and date press Start, Control Panel, Date Time Language and 
Regional Options, Date and Time
e. Before replacing the A5 Disk Drive, see the “Calibration Data Backup and Rest36ore” section in this 
chapter for details.
f. See Chapter 17, “Instrument Software” for instructions on updating the instrument software to the 
latest version.
Table 15-1  Post-Repair Testing Requirements  (Continued)
Assembly
Adjustments to Perform 
(in the order listed)
a
Performance Verifications to Perform After 
Adjustments
a