602
Functional Tests
Test Equipment
Cables
All 3.5 mm (1 meter) Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
11500E 8120-4921
All Cable, BNC (3 required) 120 cm (48 in.) BNC cable 10503A
BBA Cable, BNC 120 cm (48 in.) BNC cable
Flatness (250 kHz to 40 MHz) < 0.5 dB
10503A
BBA Cable, BNC BNC male
23 cm (9 in) max
10502A
Signal Source
All Synthesized Sweeper Frequency: 10 MHz to 26.5 GHz
Harmonic level:< - 40 dBc
Amplitude range: 10 to - 20 dBm
Frequency Accuracy: 0.02%
PSG
a
83630B,
83640B,
83650B
BBA Signal Generator Frequency: 250 kHz to 40 MHz
Amplitude: ~–10 dBm
Flatness: < ±0.75 dB
E4438C 33250A
Power Meter
All Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor
E4418B E4419B
All Power Sensor Frequency Range: 50 MHz to
3.66 GHz
Amplitude Range: -70 to –10 dBm
8481D 8487D
8485D
P03 Low Power Sensor Frequency Range: 50 MHz to 3.6 GHz
Amplitude Range: –70 to –10 dBm
8481D 8485D
8487D
Pxx other
than P03
Low Power Sensor Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: –70 to –10 dBm
8485D 8487D
Oscilloscope
All Keysight Infiniium Oscilloscope Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
54800B
Table 16-1
Analyzer
Option
Item Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model