EasyManua.ls Logo

Keysight N9020A - Page 632

Keysight N9020A
642 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
632
Functional Tests
BBIQ Input Frequency Response (Option BBA only)
19.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 16-10 for the I Input at 250 kHz.
20.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 1, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
a. Press Peak Search
b. Record the marker amplitude reading in Table 16-10 for the I Input
at the current RF Signal Generator frequency
NOTE Since the input is set for I + Q and the input signal is being applied to I only,
the analyzer will measure 6 dB lower than the signal generator output; in this
case
-10 dB - +6 dB = -16 dB
This will be the case for I, I, Q, and Q inputs.
This is because the Fourier transform of
Asin (w+) = Am(W-w) + 1/2 Am(W+w).
Since a signal is placed at w or -w, the result will be 20log(1/2) lower.
The transform is in voltage and not power.
21.Connect the BNC Cable between the RF Signal Generator RF Output and
the Q input on the X-Series analyzer, see Figure 16-10
22.Connect the 50 Ohm termination to the Q input on the X-Series analyzer,
see Figure 16-10
23.Set the RF Signal Generator frequency to 250 kHz.
24.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 16-11 for the Q Input at 250 kHz.
25.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 16-11, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
Press Peak Search
Record the marker amplitude reading in Table 16-11 for the Q Input at the
current RF Signal Generator frequency

Table of Contents

Other manuals for Keysight N9020A

Related product manuals