634
Functional Tests
BBIQ Input Frequency Response (Option BBA only)
31.In Table 16-10 and Table 16-11, calculate the Frequency Response for
each input by subtracting the Marker Amplitude at 250 kHz for that input
from the Marker Amplitude for each frequency greater than 250 kHz.
32.Compare the Frequency Response values to the Test Limits.
Frequency Response = Marker Amplitude (Freq > 250 kHz) – Marker Amplitude (250 kHz)