66 
Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
AIF Step Gain Algorithm
Measures the relative gain of the IF for high gain (10 dB) 
versus the low gain (0 dB) paths. In normal instrument 
operation these two states can be controlled by the user 
interface key called “Swept IF Gain”.
Uses the Comb calibrator on the A-IF
A2 Analog IF  
AIF 140 Variable Gain Alignment Algorithm
(Option B85, B1A, or B1X)
A25 Wideband A-IF  
LO Power Alignment Algorithm
Measure max LO power then backs off for each band
A13 Front End 
Assembly or A15 Front 
End Controller
 
AIF Variable Attenuator Algorithm
Measures the gain in all settings of the A-IF 1 dB step 
attenuator. A reference power is measured with atten 
4dB
The Comb calibrator on A-IF is used to provide a CW 
signal.
A2 Analog IF  
Mechanical Attenuator Algorithm
Measures the attenuation of all DC coupled, and one AC 
coupled (10 dB) settings of the RF input attenuators.
Reference atten setting is 10 dB
A9 and A10 RF Input 
Attenuators
50 MHz calibrator on A16 
Reference
Electrical Attenuator Algorithm
Each setting from 0 to 24 dB in steps of 1 dB is 
measured. The reference condition is electrical 
attenuator set to 0 dB and RF input attenuators set to 10 
dB. (Option EA3)
A13 Front End 50 MHz calibrator on A16 
Reference.
Electrical Attenuator Tail Algorithm 
Attenuation set to 0 dB (Option EA3)
A13 Front End  
AIF 140 Variable Attenuator Algorithm 
Ref power at atten 12 dB (Option B85, B1A, or B1X)
A25 Wide band A-IF  
DIF 40 Variable Attenuator Algorithm
Ref power at atten 21 dB (Option B40)
A3 Digital IF
Table 2-2  Initial Alignments
Alignment Description Most Probable 
Hard ware Failure
Related Hardware