87
Instrument Messages
Event Messages
–310 System Error; No 
supported source 
found
Signal source at given IP address is not responding / IP does not belong to a 
source. Check IP address and network connection
–310 System Error; source 
connection lost, 
check interface 
connection
Signal source at given IP address is not responding / IP does not belong to a 
source. Check IP address and network connection
–310 System error; The 
configured PTP 
hardware driver could 
not be instantiated.
The PTP driver failed on initialization
–310 System error; The PTP 
hardware driver 
reported a 
configuration error.
Failure in the execution of the PTP driver. The most likely cause of this error is a 
mismatch between versions of the PTP driver and the LXI middleware
–310 System error; The PTP 
ordinary clock 
reported a 
configuration error.
Failure in execution of the DMC libraries’ PTP controller
–310 System error; The 
Trigger alarm delayed 
LAN event was not 
scheduled due to an 
existing pending 
event.
Delayed LAN events cannot occur too close together (within 20 ms)
–310 System error; The 
Trigger alarm delayed 
LAN event was not 
scheduled due to a 
conflict with an 
existing scheduled 
alarm.
Delayed LAN events cannot occur too close to a scheduled Alarm (within 20 
ms)
–310 System error; The 
Trigger alarm was not 
scheduled due to a 
conflict with an 
existing scheduled 
alarm.
Alarms cannot be scheduled to happen too close together (within 20 ms)
–311 Memory error There is a physical problem with the instrument memory, such as a parity error
–312 PUD memory lost Protected user data saved by the *PUD command has been lost
–313 Calibration memory 
lost
The nonvolatile calibration data used by the *CAL? command has been lost