Instrument Messages
Event Messages
–221 Settings conflict;
Signal Track is
turned off when Zero
Span is selected
Signal Track is not available when you have selected Zero Span. So if Zero
Span is entered while in Signal Track is On, Signal Track is turned off
–221 Settings Conflict;
Span limited to XXX
–221 Settings conflict;
Span Zoom is not
available in Zero
Span
Span Zoom does not work with a time domain X-axis. You must select a span
greater then 0 Hz
–221 Settings conflict;
Step keys are not
available to modify
this function
You should select a specific value for this function. Using the
Up/Down
step keys to scroll through values is not allowed
–221 Settings conflict;
Sweep Setup only
available in swept
measurements
The current measurement uses FFT mode, so does not use the Sweep Setup
menu
–221 Settings conflict;
Sweep Time cannot be
auto-coupled in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations. So sweep time
settings cannot be adjusted
–221 Settings conflict;
Sweep Time cannot be
auto-coupled while in
Zero Span
You cannot send the remote command to set the sweep time to auto while
you are in zero span
–221 Settings conflict;
Sweep Time cannot be
set while in FFT
sweeps
The sweep time for FFT sweeps is set by the calculations. It cannot be
manually controlled
–221 Settings conflict;
Swept IF Gain High
not available when
FFT IF Gain = Manual
Low
When FFT IF Gain is manually set to Low, you cannot set the Swept IF Gain to
High because that would make the Reference Level couplings wrong in
swept mode
–221 Settings conflict;
Swept LO not
available when freq
mode is Fixed
SCPI only message. The LO Mode cannot be set to Swept when the freq
mode is set to fixed. Change the freq mode away from fixed, or perform the
measurement at several fixed frequencies
–221 Settings conflict;
Swept Type=Swept is
not available while
in Gated FFT
If you have selected gated FFT then you are using the FFT sweep type and
you cannot select the swept type of sweeping