3 Short-Range Comms & IoT Mode
3.8 SEM Measurement
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier?
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier:COUPle ON | OFF | 1 | 0,
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:STOP:RCARrier:COUPle?
Example
:SEM:OFFS:INN:LIST:STOP:RCAR -30, -30, -30, -30, -30, -30
:SEM:OFFS:INN:LIST:STOP:RCAR?
:SEM:OFFS:INN:LIST:STOP:RCAR:COUP ON, ON, ON, ON, ON, ON
:SEM:OFFS:INN:LIST:STOP:RCAR:COUP?
Notes Comma-separated list of values
OFFSet 1 is for BTS, 2 for MS. Default is BTS
Couplings Coupled to Rel Start if “Auto” is selected, that is, Start is made the same as Stop
Preset 0dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB | 0dB, 0 dB, 0 dB, 0 dB, 0 dB, 0
dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB, 0 dB
ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON, ON | ON, ON, ON, ON, ON, ON, ON,
ON, ON, ON, ON, ON
State Saved Saved in instrument state
Saved in instrument state
Range Auto|Man
Min/Max -200 dB/50 dB
Fail Mask
Selects one of the logics for fail conditions between the measurement results and
the test limits:
ABSolute and RELative both check the results against the respective limit
OR checks against both limits, failing if either of the limits is broken
AND only displays a fail if both of the limits are broken
The absolute or relative power limit value for each inner offset channel can be set
remotely with [:SENSe]:SEMask:OFFSet[n]:INNer:LIST:ABSolute or
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:RCARrier.
You can turn off (not use) specific inner offset channels remotely with
[:SENSe]:SEMask:OFFSet[n]:INNer:LIST:STATe.
When sending the remote command, missing values are not permitted; that is, if you
want to change values 2 and 6, you must send all values up to 6. Subsequent values
remain unchanged. The query for this parameter always returns 12 values.
Remote
Command
[:SENSe]:SEMask:OFFSet[1]|2:INNer:LIST:TEST ABSolute | AND | OR | RELative,