3 Short-Range Comms & IoT Mode
3.2 Modulation Analysis Measurement
via SCPI, but is not changeable. In such applications the control is grayed-out. Attempts to change the
value via SCPI are ignored, but no error message is generated
Dependencies Does not appear in Spectrum Analyzer Mode
Preset
VXT Models M9410A/11A 0dB
All Others 10dB
State Saved Saved in instrument state
Min 0dB
Max
VXT Models M9410A/11A 50dB
All Others 20dB
Mixer Lvl Offset
This is an advanced setting to adjust target Range at the input mixer, which in turn
affects the signal level in the instrument’s IF. This setting can be used when
additional optimization is needed after setting "Peak-to-Average Ratio" on page
1653. Positive values of offset optimize noise performance over distortion, negative
values optimize distortion performance over noise.
Remote Command
[:SENSe]:POWer[:RF]:RANGe:MIXer:OFFSet <real>
[:SENSe]:POWer[:RF]:RANGe:MIXer:OFFSet?
Example
:POW:RANG:MIX:OFFS -5 dB
Preset 0dB
State Saved Saved in instrument state
Min
VXT Models M9410A/11A -34dB
All Others -35dB
Max 30dB
3.2.3.5 Signal Path
Contains controls that pertain to the routing of the signal through the frontend of
the instrument.
In general, only appears in instruments whose hardware supports this signal
routing. For example, this tab does not appear in many of the modular instrument
products, including VXT Model M9420A, or UXM.
Short Range Comms & IoT Mode User's &Programmer's Reference 225