Loading...
3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
The RF spectrum and AF spectrum traces are not averaged.
The demod waveform window will display only the demod trace.
The frequency drift window will display only the frequency drift trace.
The metrics window will display only a “current” column.
Metrics will display to two significant digits.
The Meas Bar will not display the Average/Hold number.
Max Hold metrics over SCPI will return SCPI not a number.
Demod min trace, demod max trace and demod avg trace will return default values
in a meas?, read?, and fetch? And when exported to a .csv file
Max Hold column shows “---”.
Demod & Analysis Setup Table
The Demod & Analysis Setup Table provides a more straightforward way to
configure the parameters.
400 Short Range Comms & IoT Mode User's &Programmer's Reference