3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Number of Segments
Sets the total number of segments in the STS field.
Remote
Command
[:SENSe]:HUWB:STS:SEGMent:NUMBer N1 | … | N4
[:SENSe]:HUWB:STS:SEGMent:NUMBer?
Example
:HUWB:STS:SEGM:NUMB N2
Couplings See "PHY Mode" on page 501
–
For ERDEV-BPRF, Segment Number is fixed at 1
–
For ERDEV-HPRF, Segment Length is selectable
Available when PHY Mode is not Non-ERDEV and "STS Packet Configuration" on page 502 is not No
STS
Preset
N1
State Saved Saved in instrument state
Range
1|…|4
Extra Gap (x4 chips)
Sets an extra gap (in units of 4 chips) between the Payload and the beginning of
STS.
Remote
Command
[:SENSe]:HUWB:STS:EGAP <integer>
[:SENSe]:HUWB:STS:EGAP?
Example
:HUWB:STS:EGAP 2
:HUWB:STS:EGAP?
Couplings Available when "PHY Mode" on page 501 is not Non-ERDEV and "STS Packet Configuration" on page
502 is After PHR/Payload (2)
Preset 0
State Saved Saved in instrument state
Min 0
Max 127
Key (128bits)
Sets a 128-bit STS key in hexadecimal format.
Remote
Command
[:SENSe]:HUWB:STS:KEY <string>
[:SENSe]:HUWB:STS:KEY?
Short Range Comms & IoT Mode User's &Programmer's Reference 509