6 Input/Output
Accesses menus that let you control the Input/Output parameters of the instrument.
In general, these are functions associated with external connections to the
instrument, either to the inputs or the outputs.
Input/output connections tend to be based on situation-specific hardware set up.
For that reason, input/output settings do not, in general, change when you perform
a Mode Preset. You can revert to the default values in one of three ways:
Use Restore Input/Output Defaults, in the Input/Output menu
Use System->Restore System Defaults->Input/Output Settings
Use System -> Restore System Defaults->All
The settings survive a Preset and a Power cycle.
A few Input/Output settings do respond to Mode Preset. For example, if the
Calibrator is on, Preset turns it off, and if DC coupling is in effect, Preset switches it
to AC. These exceptions are noted in the SCPI tables for the excepted functions.
Input/Output features are common across multiple Modes and Measurements. In
general, they do not change when you change Mode or Measurement, although
some controls appear only in certain measurements.