3 Short-Range Comms & IoT Mode
3.9 Occupied BW Measurement
Trace Type Query Returns:
WRITe NORMal
AVERage AVERage
MAXHold POSitive
MINHold NEGative
Preset
AVERage
State Saved Saved in instrument state
Range
NORMal|AVERage|POSitive|SAMPle|NEGative|RMS
Detector Select Auto/Man
Sets the Detector mode to Auto (ON) or Man (OFF). In Auto, the proper detector is
chosen based on rules that take into account the measurement settings and other
instrument settings.
When you select any "Detector" on page 1349 manually, this setting reverts
automatically to Man (manual).
Remote
Command
[:SENSe]:OBWidth:DETector:AUTO ON | OFF | 1 | 0
[:SENSe]:OBWidth:DETector:AUTO?
Example
:OBW:DET:AUTO ON
:OBW:DET:AUTO?
Couplings When ON, the query "Detector" on page 1349 returns values that depend on the setting of "Trace Type"
on page 1731 as follows:
Trace Type Query Returns:
WRITe NORMal
AVERage AVERage
MAXHold POSitive
MINHold NEGative
Preset
ON
State Saved Yes
3.9.10.5 Trace Function
Contains controls to:
Copy and Exchange traces
Preset or Clear all traces
Short Range Comms & IoT Mode User's &Programmer's Reference 1351