Keysight N9927-90001 User’s Guide 329
Pulse Measurements Mode – Option 330
Pulse Measurement Settings
Then choose from the following:
— Target
, Peak, and Minimum. Learn more about these standard features
beginning in “Searching with Markers” on page 452.
— Rise Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value BEFORE the Peak.
— Fall Time
Select, then set a value in dB. Delta markers are created and
placed at the Peak and at the specified value AFTER the Peak.
Auto Analysis
Auto Analysis reports the following TEN pulse measurements:
Rise and Fall time in Marker Search uses different criteria than in Auto
Analysis and will therefore result in different reported values.
1 - Average Power Average power level of all measured data
points
2 - Peak Power Highest reading of all measured data
points
3 - Peak minus Average Difference between the above two results
4 - Pulse Top Power value based on the specified Pulse
Top setting.
5 - Rise Time The time it takes for a pulse to rise from
10% to 90% of its peak value.
6 - Fall Time The time it takes for a pulse to fall from
90% to 10% of its peak value.
7 - Pulse Width The time between the rising edge and
falling edge of a pulse.
8 - Pulse Period
The time of one complete pulse cycle
9 - Duty Cycle The ratio of the pulse width / pulse period.
10 - Pulse Rep Freq The rate, in pulses / sec, at which pulses
recur.