EasyManua.ls Logo

Keysight N9913A - How to Make Dtf Measurements; Chapter 4 DTF (Distance to Fault) Measurements How to Make DTF Measurements

Keysight N9913A
566 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58 Keysight N9927-90001 User’s Guide
DTF (Distance to Fault) Measurements
How to make DTF Measurements
Optional settings
How to make DTF Measurements
Before starting, you may need the following:
Jumper cable or adapter to connect the beginning of the DUT to the
FieldFox.
LOAD with correct connector type and gender to terminate the end of the
DUT (if possible).
The known length and cable type of the DUT. If the cable type is not known,
then the Cable Loss (dB/Meter) and Velocity Factor of the DUT are required.
1. Connect any necessary jumper cable or adapter to the FieldFox RF OUT
port. Do NOT connect the DUT.
2. Press Preset
then Preset to return the FieldFox to the default settings.
3. Then Mode
then CAT.
4. Then Meas 1
then DTF.
5. Press Freq/Dist
, then Stop Distance and enter the length of the DUT. You
can optionally set the Start Distance
6. Press Cal 5
and follow the Cal prompts. Learn all about Calibration in
“How to Perform a Calibration” on page 121.
7. Disconnect any components or antenna that should NOT be measured and
connect a LOAD at the end of the DUT.
8. Press Meas Setup 4
then DTF Cable Specifications.
“Coupled Frequency” on page 62
“Cable (Correction) Specifications” on page 63
“Window Settings” on page 66
“DTF Units” on page 67
“Calculated DTF values” on page 67
“About Alias Faults” on page 67
“All about Markers” on page 446
“All about Limit Lines” on page 461
“Saving and Recalling Files” on page 506
Trace Math is NOT available in DTF Measurements.

Table of Contents