364 Keysight N9927-90001 User’s Guide
AM/FM Metrics (SA Only and Option 355)
SINAD and THD Calculations
SINAD and THD Calculations
FieldFox converts the audio waveform to the audio frequency spectrum
using an FFT (i.e., The FFT result is not displayed). Refer to Figure 21-4 on
page 364.
The FFT waveform is used to calculate SINAD and THD distortion metrics
The fundamental signal, harmonics, and noise from 0 Hz to 24 kHz are
applied to the formulas. Refer to Figure 21-5 on page 365.
For more on calculations, refer to “Overview of Measurement Types” on
page 357.
See also, http://literature.cdn.keysight.com/litweb/pdf/5991-4913EN.pdf
Figure 21-4 Audio Waveform with FFT Calculations

Table of Contents