Keysight N9927-90001 User’s Guide                                                            435
Noise Figure Mode (Option 356 – CPU2 Only)
Calibration Wizard
— When the DUT output noise power is relatively low (i.e., the gain plus noise 
figure of the DUT is less than 15 dB), measurement accuracy may also be 
improved by using a low loss isolator on the preamplifier input. Due to the 
output reflection coefficient of the DUT, reverse noise waves from the 
preamplifier, when present, will contribute a different amount of noise 
power during the measurement of the DUT than during the user calibration. 
And, this change in receiver noise power can impact the measurement 
accuracy.
— It is not necessary to use the same noise source in the user calibration as 
will be used in the measurement. In some cases, it might be better to use a 
different noise source in the user calibration than will be used in the 
measurement. If this is done, it is important to set the ENR table for the 
calibration noise source prior to the calibration, and to set the measurement 
noise source prior to measuring the DUT.
— Prior to starting a user calibration, it is a good practice to connect the 
preamplifier and noise source that will be used during the user calibration 
and perform a noise figure measurement of the preamplifier itself (without a 
user calibration). Use the same bandwidth and frequency settings that will 
be used in the user calibration. A successful noise figure measurement 
ensures that all of the connections (bias, control, and RF) have been made 
properly. The user calibration can be started immediately following the 
successful noise figure measurement of the preamplifier, with no change to 
the setup.