Keysight N9927-90001 User’s Guide 493
System Settings
System Configuration
Hold-over mode
When the GPS signal is lost, the last GPS reading is used as a hold-over
correction value for the internal reference. The hold-over correction value is no
longer used, but instead the factory-set correction value is used, when any of
the following occur:
when GPS is disabled.
when Reference Source is set to Internal.
when the FieldFox resumes from Standby mode or shut down.
Frequency Reference Status
When switched to Ext or GPS, one of the following annotations appears in the
lower-left corner of the screen:
EXT R – External reference is selected.
GPS R – GPS reference is selected.
GPS H – Internal Reference is used with last ‘hold-over’ correction reading.
If the FieldFox does NOT find an appropriate Ext or GPS reference signal, then
Ext Ref UNLOCKED is displayed in the lower-left corner of the graticule area.
This can occur when any of the following conditions exist:
NO reference signal is present.
A low or very high-level external reference signal is present. For best
results, the signal should be between –5 dBm and +5 dBm.
The external reference is off-frequency.
To learn the expected measurement accuracy when connected to a frequency
reference source, see the FieldFox specifications in “Specifications/Data
Sheet” on page 549.
Reference OUT
The internal 10 MHz frequency reference is available from the Ref/Trig output
connector on the right-side panel. Use this signal as a common frequency
reference when making measurements using external equipment. See this
connector in “Top Panel” on page 36.