EasyManua.ls Logo

Keysight PNA-X Series - Sensitivity

Keysight PNA-X Series
47 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 27
Find us at www.keysight.com
Overview S93029B/A without Option 029 S93029B/A with Option 029
NF measurements, converters
Max DUT input frequency (GHz) 8.5/13.5/26.5/43.5/50/67 8.5/13.5/26.5/43.5/50/67
Max DUT output frequency, using
low-noise receivers (GHz)
N/A 8.5/13.5/26.5/43.5/50
Max DUT output frequency, using
standard receivers (GHz)
1
8.5/13.5/26.5/43.5/50/67 8.5/13.5/26.5/43.5/50/67
Calibration accessories for low-noise
receivers
N/A Cal kit or ECal, power meter, 346C or 346C-K01 noise
source (optional)
Calibration accessories for standard
receivers
Cal kit or ECal, power meter Cal kit or ECal, power meter
PNA Family Applications and Options (Continued)
Noise figure measurement summary
1. External preamplifier and filter(s) required if DUT excess noise (gain plus noise figure) is < 30 dB up to
20 GHz, < 40 dB up to 50 GHz, or < 45 dB up to 67 GHz.

Other manuals for Keysight PNA-X Series

Related product manuals