6 Calibration Procedures
Adjustments procedures
216 U3606A User’s and Service Guide
2 Press  Shift > EXT to enable remote sensing. When the U3606A is 
operating in remote sensing mode, the EXT annunciator on the front 
panel is illuminated.
3 Perform  step 2 through step 7 of the “CC output adjustment procedure 
— front output terminals” on page 211. Refer to Table 6- 11 on page 212 
for the constant current output adjustment items.
4 Repeat  step 3 through step 7 of the “CC output adjustment procedure — 
front output terminals” on page 211 for the S1 (30 V/1 A) range. Press 
Voltage to select the S1 range (the S1 annunciator is illuminated).
For adjustment item “LOAd”, connect a 30 Ω, 50 W load across the 
front panel + rear output terminal and the I (red) terminal. Leave the 
– rear terminal and LO (black) terminal connected.
Do not remove the short bar between the rear panel sense terminals (+S 
and –S) are the rear panel output (+ and –) terminals. See “Remote 
sensing connections” on page 100 for more information on how to 
connect the load leads.