U9361C/F/G/M RCal Measurement Guide 17
Getting Started
RCal Software – Calibration and Application
This message will be displayed during the calibration process:
When calibration is complete, you should see the following message which will
disappear automatically:
To cancel a calibration in progress, press the Cancel button or use the Esc key.
After calibration, you should see a time stamp for all calibrated rows under the
Last Calibrated column.
Applying a Calibration
To apply the correction, select the Apply column in the row. The instrument will
compare its current state to all corrections in the active RCal correction pool
and apply correction factors according to the Match State rules. The Applied
column should now show Yes, indicating that the selected calibration is being
applied.
When calibrations are active, the Corr CCorr (Complex Correction) option on
the measurement panel should show "RCal" in orange as shown below.