20 U9361C/F/G/M RCal Measurement Guide
Getting Started
RCal Software – Calibration and Application
Situations that are in between will utilize interpolation to get the desired
correction.
Multiple Rows, Overlap, and Dual Corrections
If you create multiple rows, calibrate them, and select Apply in more than one
row, the RCal system may raise apply-time warnings. As long as one row
matches the current state, the calibration will be applied successfully for that
row despite the warning.
If two or more rows both match state, they are considered overlapping and
yield a double-correction. This overlap may be over a portion of the spectrum
trace. This usually is undesired, and should be avoided or fixed by selecting
Apply in a single row, or fewer rows, until the warning goes away, or
configuring RCal to avoid overlap.
The following warning indicates that more than two rows match state:
When performing a calibration, no corrections will be made outside of the
calibration bandwidth. However, if performing a calibration on a signal with a
wider bandwidth than the calibration bandwidth, corrections will still be
applied within the calibrated bandwidth.