U9361C/F/G/M RCal Measurement Guide 33
Making Measurements
Performing a Simple Magnitude Calibration
12. Disconnect the cable from the RCal RF Out port and connect it to the
signal generator RF Out port.
13. Select Trace. In the Select Trace drop-down, select Trace 2 to verify the
calibration by adding another trace.
14. Set Trace Type to Max Hold, and then in the View/Blank area, select
Active.
The trace will begin to flatten as the maximum points are saved. Notice
the improvement in the measurement.