36 U9361C/F/G/M RCal Measurement Guide
Making Measurements
Measuring Magnitude Cal with Multiple States
6. Select Meas Setup> Source tab > Source Mode> Tracking to set up the
source control to sweep across the frequency range.
7. Select Meas Setup > Source > RF Output > On to enable the RF output of
the signal generator.
8. Select Trace > Select Trace > Trace 1, set Trace Type to Max Hold, and set
View to Active.
9. Once the trace has stabilized, save it for comparison by setting
View/Blank to View.
10. Select Meas Setup > Source tab> RF Output > Off to disable the RF
output on the signal generator.
11. Disconnect the cable from the signal generator and connect it to the RCal
module.