70 PMC-A
・ If the output voltage is being controlled using an external
voltage (CV-V mode) on Type II models, the output volt-
age at full scale cannot be adjusted using the variable
resistor. Adjust the external voltage to obtain the full
scale voltage.
8.
Set the output voltage to the maximum output volt-
age.
When using local control, turn the VOLTAGE knob
clockwise all the way. When using analog remote control,
set the control signal to 10 V or 10 k
Ω.
9.
Adjust Vout MAX so that the output voltage (DVM
reading) is slightly higher than the rated voltage.
Both Type I and Type II: Variable resistor A
■ Voltmeter at full scale
10.
Set the output voltage (DVM reading) to the rated
voltage.
11.
Adjust V METER FS so that the voltmeter reading
on the power supply is equal to the DVM reading.
Type I: Variable resistor 1
Type II: Variable resistor 2
■ Preset voltage indicator at full scale
12.
Adjust V LIMIT FS so that the voltmeter reading on
the power supply is equal to the external DVM read
-
ing when the LIMIT switch is pressed while delivering
rated voltage.
Both Type I and Type II: Variable resistor 4