7-8 TOS5051A/5050A
MODE?
Queries the timer setting, the lower cutoff current setting, the voltage mode setting,
and the voltage range setting.
Response
Returned as in the example below.
TIMEON,LOWEON,VOLTAC,RANG2.5
TIMEOFF,LOWEOFF,VOLTDC,RANG5.0
MEASURE? (MEAS?)
Queries the measured value of the tester.
Response
The response varies when the test is in progress and when it is not.
When the test is in progress
Returns the measured voltage, measured current, and actual test time.
When the test is not in progress
Returns the result of the previous test as follows:
Returns the maximum measured voltage, maximum measured current,
actual test time, and status (any of the following).
<PROTECT>, <PASS>, <U_FAIL>, <L_FAIL>, <STOP>, or <P_ON>
(<P_ON> indicates the response against the MEASURE? command
immediately after the power is turned ON (condition in which the test is
not performed even once).)
TMODE (TMOD)
Sets the talk mode.
Command examples
Response
Returns OK.
The talk mode setting specied by a command is reset to the panel setting when
the tester is power cycled.
TMOD 0
:
Set to talk mode 0.
TMODE 1:
Set to talk mode 1.
Tmode 2:
Set to talk mode 2.