TOS5051A/5050A 7-5
Chap.7
RS-232C Interface
If nothing appears on the uorescent display after one minute elapses, redo the
procedure from the beginning.
4.
Check that the talk mode number specied in step 2 is displayed.
If you carry out initial setup of the data, the talk mode is set to 0. For details, see
section 5.1, “Initial Setup” (Page 5-2 ).
7.5 Before Using the RS-232C
Maximum measured voltage and measured voltage
The voltmeter on the front panel displays the upper 3 digits, but the RS-232C
returns 4 digits.
(Example)
Maximum measured current and measured current
If the test ends due to UPPER FAIL or LOWER FAIL, the returned current is
not the measured value, but the upper cutoff current or the lower cutoff current.
Actual test time
The maximum actual test time that can be returned is 999 seconds.999 seconds
is returned also on tests lasting longer than 1000 with the timer function turned
OFF.
When the tester is set to special test modes
When DOUBLE ACTION or MOMENTARY is turned ON using the TEST
MODE switch on the rear panel, the test will not start even when a START com-
mand is sent to the tester. Turn OFF DOUBLE ACTION and MOMENTARY.
Response Panel display
4991 V 4.99 kV
4999 V 4.99 kV
PROTECTION
READY
TEST PASS
AC
kV
mA
FAIL
DC
s
CURRENT TIME
UPPER LOWER
KEYLOCK
REMOTE
TIMER ON
LOWER ON
UPPER
LOWER
RANGE
kV
VOLTAGE
Ta
lk
mo
d
e
num
b
er
Version number
Model name 3-digit display
TOS5051A 51A
TOS5050A 50A
NOTE