KINETEC CEM™ 17/22 EN
USING THE KINETEC PATIENT PAD KIT
The device KINETEC CEM™ is delivered with 3 straps:
• 1 - Arm strap (A)
• 2 - Forearm straps (B)
• 1 - Elbow protection (C)
All these straps are used the same way (see
pictures).
Do not tighten the straps too much.
FOR OPTIMAL HYGIENE,
A NEW SET OF PADS SHOULD BE USED FOR EACH PATIENT.
Recommendations to obtain maximum hygiene for the straps:
• Sterilizing the straps (if necessary): Sterilize at 134 °C for 18 minutes.
• Disinfecting the straps:
Wash at 30°C, using a disinfectant solution during the rinse cycle.
Example of product which can be used: Solution "Baclinge" at 0.125 % or "Souplanios" at 0.125% from
ANIOS Laboratory. A complete list of distributors in your country is available on request.
A
B
C