KINETEC CEM™
22/22
EN
PRODUCT INFORMATION
Guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration - Electromagnetic immunity
The device KINETEC CEM™ is intended for use in the electromagnetic environment specified below. The customer or the user of the device
KINETEC CEM™ should ensure that it is used in such an environment.
Immunity test
Test level according to
IEC 60601
Compliance level
Electromagnetic environment –
Guidance
Mobile and portable RF communication devices should not be used
closer to any part of the device KINETEC CEM™, including its
cables, than the recommended separation distance, calculated
based on the equation applicable to the emitter's frequency.
Recommended separation distance
Conducted RF interference
IEC 61000-4-6
3 Veff
from 150 kHz to 80 MHz
3 V
d = 1.2
P
Radiated RF interference
IEC 61000-4-3
3 V/m
from 80 MHz to 2.5 GHz
3 V/m
d = 1.2
P
from 80 MHz to 800 MHz
d = 2.3
P
from 800 MHz to 2.5 GHz
where P is the emitter's maximum output power characteristic in
watts (W), according to the emitter's manufacturer, and d is the
recommended separation distance in metres (m).
The field intensities of fixed RF emitters, determined by an on-site
electromagnetic investigation
a
, should be below the compliance
level in each frequency range
b
.
There may be interference near appliances bearing the following
symbol:
NOTE 1 At 80 and 800 MHz, the highest frequency range is applicable.
NOTE 2 These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection by structures,
objects and people.
a
The field intensity of fixed emitters such as base stations for radio-telephones (cellular/cordless) and land mobile radios, amateur radio, AM/FM
radio broadcasts and TV broadcasts cannot be predicted exactly in theory. To evaluate the electromagnetic environment due to fixed RF
emitters, an on-site electromagnetic investigation should be considered. If the field intensity measured where the device KINETEC CEM™ is
used exceeds the aforementioned applicable RF compliance level, the device KINETEC CEM™ should be monitored to check that it is working
normally. If abnormal results are observed, additional measures may be necessary, such as reorienting or repositioning the device KINETEC
CEM™.
b
Over the frequency range 150 kHz to 80MHz, field intensities should be less than 3V/m.
Recommended separation distances between mobile and portable RF communication devices and the
KINETEC CEM™
The device KINETEC CEM™ is designed to be used in an electromagnetic environment in which radiated RF interference is controlled. The
customer or user of the device KINETEC CEM™ can help prevent electromagnetic interference by maintaining a minimum distance between mobile
and portable RF communication devices (emitters) and the device KINETEC CEM™, as recommended below, according to the communication
device's maximum output power.
Maximum assigned output power
for the emitter
W
Separation distance according to the emitter's frequency
m
from 150 kHz to 80 MHz
d = 1.2
P
80 MHz to 800 MHz
d = 1.2
P
from 800 MHz to 2.5 GHz
d = 2.3
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For emitters whose assigned maximum emitted power is not given above, the recommended separation distance d in metres (m) can be estimated
using the equation applicable to the emitter frequency, where P is the emitter's maximum emission power characteristic in watts (W), according to the
latter's manufacturer.
NOTE 1 At 80 and 800 MHz, the separation distance for the highest frequency range is applicable.
NOTE 2 These directives cannot be applied in every situation. Electromagnetic propagation is affected by absorption and reflection by