Conguration: Overview
Conguration (default in bold print)
Current output 1 Cond
RANGE 4 ... 20 mA / 0 ... 20 mA
CHANNEL COND / TMP
(Current output curve)
LIN / BILIN / LOG (LOG for S/cm and S/m only)
LIN BEGIN 0/4 mA
Floating-point input, 000.0 mS/c
END 20 mA
Floating-point input, 100.0 mS/c
LOG BEGIN 0/4 mA
Selectable decades:
END 20 mA
Decades (see above)
BI LIN BEGIN 0/4 mA
Range depending on selected channel
END 20 mA
Range depending on selected channel
CORNER X
Conditions for bilinear characteristic:
Vertex X: BEGIN ≤ CORNER X ≤ END (rising)
BEGIN ≥ CORNER X ≥ END (falling)
CORNER Y
Default: 12 mA
Vertex Y: (0 mA) 4 mA ≤ CORNER Y ≤ 20 mA
TMP
°C
BEGIN 0/4 mA –50 ... 250 °C (000.0 °C)
END 20 mA –50 ... 250 °C (100.0 °C)
TMP
°F
BEGIN 0/4 mA –58 ... 482 °F (032.0 °F)
END 20 mA –58 ... 482 °F (212.0 °F)
ERROR
Output current for error message OFF / FAIL / FACE (Sensoface)
FILTERTIME
Output filter, time interval 0 s ... 120 s (0 s = filter OFF)
HOLD
Output current for HOLD LAST / FIX
Current output 2 Default setting CHANNEL: TMP
(other settings like OT1)
Correction Cond
OFF / LIN / NLF / compensation for ultrapure water: NaCl, HCl,
NH3, NaOH
LIN TC LIQUID 00.00 ... 19.99 %/K (00.00 %/K)
REF TEMP 000.0 ... 199.9 °C (025.0 °C)