83
Sensocheck, Sensoface Sensor Monitoring
Sensocheck continuously monitors the sensor and its wiring.
The three Sensoface indicators provide information on required mainte-
nance of the sensor. Additional icons refer to the error cause.
Pressing the info key shows an information text.
Note:
The worsening of a Sensoface criterion leads to the devaluation of the Sensoface
indicator (Smiley gets “sad”). An improvement of the Sensoface indicator can only
take place after calibration or removal of the sensor defect.
Sensoface message
The Sensocheck message is also output as error message Err 15.
The alarm contact is active, the display backlighting turns red,
output current OUT is set to 22 mA (when congured correspondingly).
All other Sensoface message can be output via a contact
(relay contacts, alarm --> “FACE”).
Disabling Sensocheck and Sensoface
Sensocheck can be switched o in the conguration menu (then Sensoface is also
disabled).
Exception:
After a calibration, a smiley is always displayed for conrmation.