Use of relays
Select process variable
Limit 1
switching characteristics (function)
Limit 1 contact type
Limit 1 setpoint
Limit 1 hysteresis
Limit 1 delay
Relay Contacts: Limit Function, Hysteresis
(Example: relay 1)
1 Press menu key.
2 Select CONF using ,
press enter.
3 Select RELAY1 menu using  keys,
press enter.
Specify relay contact function: LIMIT.
4 All items of this menu group are indicated by the
LM1:“code.
Press enter to select menu,
edit using arrow keys (see next page).
Confirm (and proceed) by pressing enter.
5 Exit: Press meas key until the [meas] mode indicator is
displayed.
Conguring the Relay Contacts