71
CondI
Zero calibration in air / with calibration solution (toroidal sensors)
Display Action Remark
Select Calibration.
Press enter to proceed.
Select CAL_ZERO
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is in
HOLD mode.
Calibration in air
Edit digits until the lower
display indicates Zero
Calibration in air (AIR-SET)
for SE680-M and Memosens
sensors: Press enter to start
AIR-SET. When zero point and
cell factor are displayed, press
enter to conrm.
The device shows the cell
factor (at 25 °C) and the zero
point.
Sensoface is active.
Use the arrow keys to select:
MEAS (exit)
REPEAT
Press enter to proceed.
Exit:
HOLD is deactivated
after a short time.