EasyManua.ls Logo

Knick Stratos MS A405 Series - Introduction

Knick Stratos MS A405 Series
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Introduction
Stratos MS is a 4-wire analyzer for use with Memosens sensors. The Model A405B
allows applications in hazardous-area Zone 2. Current is provided through a
universal power supply 80...230VAC, 45...65Hz / 24...60VDC.
The analyzer provides two 0 (4) .... 20mA current outputs for transmission of
measured value and temperature, for example.
Two oating relay contacts are available for free conguration.
You can select one of the following measuring functions:
pH
ORP
Dissolved oxygen
Conductivity measurement (conductive/inductive)
Enclosure and mounting possibilities
The sturdy molded enclosure is rated IP 67/NEMA 4X outdoor.
Material of front unit: PBT, rear unit: PC.
Dimensions: H 148 mm, W 148 mm, D 117 mm.
It is provided with knockouts to allow:
panel mounting (138 mm x 138 mm cutout to DIN 43700)
wall mounting (with sealing plugs to seal the enclosure)
post/pipe mounting (dia. 40 … 60 mm, 30 … 45 mm)
Protective hood (accessory)
The protective hood provides additional protection against direct weather exposure
and mechanical damage (available as accessory).
Connection of sensors, cable glands
For connecting the cables, the enclosure provides
3 knockouts for cable glands M20x1.5
2 knockouts for NPT 1/2” or rigid metallic conduit
Memosens sensors and connecting cables
Please visit our website for more information on our product range: www.knick.de.