51
Menu item Action Choices
Use of relays
Select in the text line using
keys:
• Limit function (LIMITS)
• Error message (ALARM)
• Rinse contact (WASH)
Press enter to confirm.
Note: The following submenu
depends on the selected setting.
Select process variable
Select desired process variable
using keys.
Press enter to confirm.
Limit 1 function
Select desired function using
arrow keys.
LoLevel: active if value falls
below setpoint
LoLevel: active if value exceeds
setpoint
Press enter to confirm.
Limit 1 icon:
Limit 1 contact response
N/O: normally open contact
N/C: normally closed contact
Select using keys.
Press enter to confirm.
Limit 1 setpoint
Enter setpoint using
keys.
Press enter to confirm.
within measuring range
Conguring the Relay Contacts