43
Conguring the Current Output
Menu item Action Choices
Time averaging lter
Enter value using  
keys.
Press enter to confirm.
Time averaging lter
To smoothen the current output, a low-pass lter with adjustable lter time constant
can be switched on. When there is a jump at the input (100 %), the output level is
at 63 % after the time interval has been reached. The time interval can be set from
0to 120 sec. If the time interval is set to 0 sec, the current output directly follows the
input.
Note:
The lter only acts on the current output, not on the display or the limit values!
During HOLD the lter is not applied. This prevents a jump at the output.
Time interval 0...120 s
1st variable
(channel 1,
e.g., Cond)
Time interval 0...120 s
2nd variable
(channel 2, e.g.,
temperature)
Display