The alarm delay time delays the color change of the display backlighting to red, the
22 mA signal (if congured), and the alarm contact switching.
Error messages can be signaled by a 22 mA output current. In addition, a relay
contact (RELAY1 / RELAY2) can be congured as alarm contact.
Menu item Action Choices
Alarm delay
Enter value using  
keys.
Press enter to confirm.
0...600 SEC
(10 SEC)
Sensocheck
Select Sensocheck (continuous
monitoring of sensor membrane
and lines).
Select ON or OFF using 
keys.
Press enter to confirm.
(At the same time, Sensoface is
activated. With OFF, Sensoface is
also switched off.)
ON/OFF
Tempcheck
To monitor the temperature
probe with TC OFF selected:
Select Tempcheck ON using 
keys.
Press enter to confirm.
Now, the temperature probe will
be monitored.
ON/OFF
Conguring the Alarm