Product Calibration
Display Action Remark
The device returns to mea-
suring mode.
From the blinking
CAL mode indicator,
you see that product
calibration has not been
terminated.
Product calibration step 2:
When the sample value has
been determined, open the
product calibration once
more
Display (3 sec)
Now the device is in
HOLD mode.
The stored value is displayed
(blinking) and can be over-
written with the lab value.
Press enter to proceed.
Display of new cell constant /
cell factor (based on 25°C).
Sensoface is active.
To exit calibration:
Select MEAS, then enter
To repeat calibration:
Select REPEAT, then
enter
End of calibration After end of calibration,
the outputs remain in
HOLD mode for a short
time.