Service
Menu item Remark
Setting the passcodes:
In the “SERVICE - CODES“ menu you can assign passcodes to DIAG,
HOLD, CAL, CONF and SERVICE modes (Service preset to 5555).
When you have lost the Service passcode, you have to request
an “Ambulance TAN“ from the manufacturer specifying the serial
number and hardware version of your device.
To enter the Ambulance TAN“, call the Service function and enter
passcode 7321. After correct input of the ambulance TAN the
device signals “PASS“ for 4 sec and resets the Service passcode to
5555.
Reset to factory settings:
In the “SERVICE - DEFAULT“ menu you can reset the device to fac-
tory settings.
NOTICE!
After a reset to factory setting the device must be reconfigured
completely, including the sensor parameters!