EasyManua.ls Logo

Kodak RVG - Customizing the Exposure Chart; Adjust the Values for Your Generator; Sensor Exposure Times

Kodak RVG
42 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjusting Exposure Time
KODAK RVG Digital Radiography Systems User’s Guide (CS4000_en) 2–7
Customizing the Exposure Chart
Table 2–1 provides you with guidelines for exposure
times for an average generator at 70kV and 8mA.
These are approximate values. Adjust the values for
your specific generator.
Adjust the values for your generator:
If the images are displayed too dark, reduce the
exposure settings.
If the images are displayed too light, increase the
exposure settings.
Add the values for corrections in the second column of
Table 2–1.
Table 2–1 Sensor Exposure
Times
Acquisition Mode High Resolution
Seconds Correction
Adult with #1 or #2
Upper incisor/canine 0.18
Upper premolar 0.24
Upper molar 0.40
Lower incisor/ canine 0.12
Lower premolar 0.18
Lower molar 0.24
Note
The exposure times
in Table 2–1 are
suggested times.
Over time and with
experience, you can
identify the settings
specific to the
configuration in your
practice.

Table of Contents

Related product manuals