A–2 Troubleshooting Images
The image is
pale and
grainy.
The exposure time is too short; increase
it. The selected acquisition mode does
not correspond to the x-ray dose used.
The generator voltage is too low
(<60 kV rms); have the generator
checked.
The generator is too far from the patient
with respect to the selected dose.
Check the monitor contrast and
brightness settings and ensure there are
no reflections on the screen.
The image is
too dark.
The exposure time is too high; lower it.
The selected acquisition mode does not
correspond to the x-ray dose used.
Check the monitor settings (contrast
and brightness) and ensure there are no
reflections on the screen
The image is
blurred.
Patient moved during exposure.
Generator head was not stable.
Use an image filter.
The image is
white.
Active face of sensor was not exposed
to x-rays.
X-ray dose is insufficient.
Sensor is not connected, or is
improperly connected.
Ensure the generator is producing
x-rays; have it checked by a certified
technician.
Table A–1 Troubleshooting