2.8 Checking Startup and Shutdown of Unit
2-25
2.8 Checking Startup and Shutdown of Unit
Turn ON the machine to check that it starts up and shuts down properly.
1 Turn ON the facility power supply.
2 Turn ON the power breaker of the unit.
3 Press the "operation" switch.
4 Check the following:
•
The message display window shows the message
"INITIALIZING" and a progress indicator bar.
• The bar extends to the right during the initialization
process.
• When the bar reaches the right end, "NETWORK
ERROR [24000]" appears in the upper part of the
message display window.
• The factory-set IP address of the machine is
displayed in the lower part of the message display
window.
5 Press the "operation" switch for at least three
seconds.
The backlight of the message display window goes off.
6 Turn OFF the power breaker.
This is the final step in the process of checking startup and shutdown of the unit.