Filling cartridges for the pressurized dosing unit page 8
KRÜSS GmbH | Borsteler Chaussee 85 | 22453 Hamburg | Germany | www.kruss.de
3 Filling cartridges for the pressurized dosing
unit
The pressurized dosing unit is used exclusively for water and diiodomethane.
1. Open the cover of the MSA. Remove the wing nut.
2. Remove the dosing cover. The cartridges are now accessible.
3. Remove the cartridge to be filled. To get an easier access to the cartridges there is a removal
tool included which is to be put underneath the lid of a cartridge.
!
The transparent cartridge is intended for water, and the brown one for diiodomethane.
Do not transpose the cartridges or their positions in the dosing unit.
4. Fit a Luer adapter to the cartridge. Fill a disposable syringe with exactly 1 mL of the test
liquid.
Diiodomethane cartridge
Water cartridge