20
3-9-3 List of Functions
For details of each function, see the following Table 3-9-3.
Table 3-9-3 Various Functions in Function Selecting Mode (1/5)
Function Name
Key Operation and WGI-400A
Movement in Function Setting Mode
Functional Description
F-01
Original Value
“……” is displayed.
Press the SET key and the input value at
this time is displayed in mV/V.
A function that displays the
raw output value of the
transducer.
F-02
Actual Load Calibration
Note: Before activating the calibration
function, always conduct digital
Zero function with no load applied
to the transducer.
A function that applies the
actual load to the transducer
and sets the indicated value
of that time.
A
pply load to the transduce
r
nput a desired numeric value (load) to
b
e indicated and press the SET key.
Set a decimal point and press the SET
k
ey. Then, the calibration value is
u
pdated.
R
emove the load.
For details, see “3-4-1
Actual Load Calibration.”
F-03
Sensitivity Registering
Calibration
Initial value:2.000 [mV/V]
4000 Display
S
et rated output of the transducer in
m
V/V and press the SET key.
I
nput the desired indicated value (rated
c
apacity) and press the SET key.
S
et a decimal point and press the SET
k
ey. Then, the calibration value is
u
pdated.
S
etting available range:
L
ess than 9999 per 0.1 mV/V
A function that conducts
calibration by registering
the rated output and rated
capacity of the transducer
without applying the actual
load.
For details, see “3-4-2
Sensitivity Registering
Calibration Function”