25
3-10 ERROR CODE TABLE
Error Display Contents Countermeasures
OFL1 Measured value is exceeding the input
range to the minus side.
Decrease the input signal.
Use within 3.2 mV/V.
OFL2 Measured value is exceeding the input
range to the plus side.
OFL3 Indicated value is below -9999. Decrease the input signal.
Re-set the sensitivity registering value.
OFL4 Indicated value is above 9999.
E-12
Sensitivity registering value is exceeding
the specified range.
Change the setting to less than 9999 per 0.1
mV/V.
E-01
Memory (RAM) failure.
Failure
Contact KYOWA or our representatives for
repair.
E-02 Memory (EEPROM) failure.