Echo Liquid Handler Software 5-31
There are two indicator lights associated with the destination plate. One
reports the current position (in/out) of the destination plate gripper stage.
The second reports the presence of a microplate loaded onto the plate
gripper.
The plate presence indicator has two states:
Clear = Plate not present;
Green = Plate present;
Flip destination plate
The Echo liquid handler can be commanded to flip its destination plate
gripper stage into the inverted or upright positions. Select the appropriate
checkbox.
If the destination plate gripper
stage is still inside the Echo
liquid handler when
commanded to flip upright, a
“Flip Dest Plate Up Error” will
appear.
The associated indicator lights report the current rotational position of the
destination plate gripper stage.
Plate dryer up/down, on/off
The Echo liquid handler can be commanded to raise or lower the plate
dryer nozzle, or to turn its vacuum supply on or off. Select the appropriate
checkboxes.
If necessary, the Echo liquid handler will move the plate dryer beneath the
source plate prior to raising it.
The associated indicator lights report the current position of the plate dryer
and whether its vacuum is turned on or off.
Anti-static bars on/off
The Echo liquid handler can be commanded to turn its anti-static bars on or
off at any time.
The associated indicator lights report whether the anti-static bars are
turned on or off.