atmospheric pressure and the output pressure should not exceed 40 in H20 (+/-
100 mbar) relative to atmospheric pressure.
For CSA (Canada) the 5000 series of gas analyzers are certified to Hazardous
Area Classification
CLASS 2258 03 - PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non-
Incendive Systems - For Hazardous Locations
Model GA 5000, GEM5000 and BIOGAS 5000 Methane Detectors; portable, battery
powered with non-field-replaceable Battery Pack P/N 20087; intrinsically safe and
providing intrinsically safe circuits (“[ib]” for Zone 1) to Model TP-5000 Temperature
Probe (Connector B) and with entity output parameters as tabulated below;
Temperature Code T1; 14 °F < Tamb. < +122°F (-10 °C < Tamb. < +50°C).
Note: This device has been investigated for electrical safety features only.
For CSA (USA) the 5000 series of gas analyzers are certified to Hazardous Area
Classification
CLASS 2258 83 - PROCESS CONTROL EQUIPMENT - Intrinsically Safe and Non-
Incendive Systems – For Hazardous Locations - CERTIFIED TO U.S. STANDARDS
Model GA 5000, GEM5000 and BIOGAS 5000 Methane Detectors; portable, battery
powered with non-field-replaceable Battery Pack P/N 20087; intrinsically safe and
providing intrinsically safe circuits (“[ib]” for Zone 1) to Model TP-5000 Temperature
Probe (Connector B) and with entity output parameters as tabulated below;
Temperature Code T1; 14 °F < Tamb. < +122°F (-10 °C < Tamb. < +50°C)
Note: This device has been investigated for electrical safety features only.
Page 10 © Copyright LANDTEC North America, Inc. V1.13
GEM5000 gas analyzer 5K-MNL-GEM5000